Technical Specifications System Function: Throughput: 200 tests/hr without ISE module Principles of measurement: Spectrophotometer, Turbidimetry Methodology: End-point, Kinetic, Two-point Sample/Reagent System Sample/Reagent tray: 24 positions for samples and 50 positions for reagents in refrigerated compartment (4-15 Sample Volume: 3-50 μl, step by 0.5 μl Reagent Volume: 25-350 μl, step by 0.5 μl Probe: Unattached probe, liquid level detection & collision protection Probe Cleaning: Automatic washing for both inside and outside, carryover <0.08% Reaction System Reaction rotor: rotating tray, containing 72 disposable cuvettes Cuvette: Optical length 6 mm Reaction volume: 180-500 μl Reaction temperature: 37±0.1 Mixer: Independent mixer Optical System: Light source: Halogen-tungsten lamp, flat field concave grating, back-dividing-light technology. Wavelength: 340 nm, 380 nm, 405 nm, 450 nm, 505 nm, 546 nm, 578 nm, 600 nm, 660 nm, 700 nm, 750 nm, 800 nm or user designed wavelengthsAbsorbance range: 0-3.0 Abs Resolution: 0.0001 Operation System:Windows XP Data Storage: Automatic storage, automatic backup, data is kept permanently Real time view of system, reagents and sample status, and temperature. Working Conditions: Power Requirements: 85-264 VAC; 50/60Hz; 900W Temperature: 10-35Humidity: 35-80% Dimension: 1000 mm (W) × 600 mm (D) × 1200 mm (H) Weight: 130 KG