Independent Working Interface for Adult and Paediatric Anesthesia Ventilator Respiration Pattern:IPPV,VCV,PCV,SIMV,SIGH,MANUAL,STANDBY,DEMO Display:10.4 TFT color screen Waveform: P-t, F-t Tidal Volume:20ml to 1600ml Respiratory frequency:1 to 65bpm IE ratio:8:1--1:8 Inspiration pause:0--50%Ti Inspiration Pressure:1--60cmH2O Trigger Pressure:-10--10cmH2O Pressure Alarm setting:High Alarm limit Set 1--60cmH2O,Low Alarm limit Set 0--20cmH2O O2 Concentration Alarm setting:High Alarm limit Set 21--99%, Low Alarm limit Set 21--50% MV Alarm setting:High Alarm limit Set 10--99L/min, Low Alarm Limit Set 0--40L/min Monitoring Parameter (under both manual and control modes): O2 concentration, MV, respiratory frequency, tidal volume, airway average pressure, PEEP, I:E ratio, air maximum pressure etc. Alarm parameters: pressure high alarm, pressure low alarm, O2 concentration high alarm, O2 concentration low alarm, MV high alarm, MV low alarm, AC power lost alarm, O2 lost alarm etc. Alarm silence function Main unit Circle absorber: compact circle absorber Bellow:0-1600ml Gauge: 4 units, one is for N2O,one is for Air, two is for O2. Flowmeterbacklight design: O2,N2O, Air five-tube flowmeter O2 flowmeter:0L/min1L/mim10L/min N2O flowmeter:0L/min1L/mim10L/min Air flow meter range:0L/min10L/min O2 and N2O Linkage Device Adjust O2 and N2O in proportion(O2 concentration is not lower than 25%) Fast for oxygen:35L/min75L/min gas source pressure O2:0.27MPa0.55MPa N2O:0.27MPa0.55Mpa Air:0.27MPa0.55MPa Anesthetic Vaporizer: two vaporizers Halothane vaporizer, Enflurane Vaporizer, Isoflurane Vaporizer, are available. Penlon vaporizer imported from UK is available.