Overview: - The machine is a high voltage leak detector and operates on a high voltage test method and the electrical conductivity of the fill product packaged in a non-conductive plastic ampoule. It is designed for fully automatic in-line testing- The machine is operational and is currently being used to leak check unit dose vials → - Currently checking 1, 2, 4 & 5ml Unit Dose Vials.- Machine inspection rate approx 1200 UDV strips per hour.- Qualification runs are performed annually.- The equipment is calibrated by BIL at routine intervals.- The machine will require appropriate maintenance and servicing.Included in Price:- Rommelag Leak Detector 921.- Conveyor In/Out.- Applicable tooling.- Operation & Maintenance documentation.Machine Specifications: Reference Data: - Machine Type: 921- Built: 1996- Machine No: 921 006Dimensions & Weight: - Weight: approx. 855kgs- Length: approx. 2625mm- Width: approx. 1670mm- Height: approx. 1870mmEmissions: - Noise: 77 dB(A)Services Specifications: Electrical: - Enclosure type: IP54- Supply Voltage: 3 x 400 V+N, PE- Frequency: 50 HZ- Current: 3 A- Control Voltage: 24 V DC- Main Fuse: 16 A- Rated Power Consumption: 2.5 kVAPneumatic: - Air Pressure: 6 bar- Max. Consumption: 80 l/minAmbient Conditions: - Temperature (machine):• Storage temperature: 15-30 ºC• Operating temperature: 15-30 ºC- Air moisture: 40-80%Current Condition: - The equipment is operational.- The machine is currently being used to leak check unit dose vials.- Currently 1, 2, 4 & 5ml Unit Dose Vials at 1200 strips per hour.