出生地 |
Guangdong China (Mainland) |
牌子的名字 |
KEYUN |
模式的数量 |
PANTHER-2 |
权力 |
电子 |
使用 |
Other |
BrACAudit Panther-2Alcohol TesterTECHNICAL DATA Details Technical DataPrinciple of measurementElectrochemical Sensor, alcohol-specificMeasurement Range (BrAC)0.000mg/L1.500mg/LResolution0.0001 mg/LAccuracy:Max. Measuring error, in relation to the ethanol standardEthanol Vapor ConcentrationC /(mg/L)DeviationC<0.400±0.020mg/L0.400≤C<1.000±5%C≥1.000±20%Repeatability of Indicating ValueEthanol Vapor ConcentrationC /(mg/L)DeviationC<0.4000.007mg/L0.400≤C<1.0001.75%C≥1.0006%CalibrationEvery 6 monthsTemperature range- operation-5 45Temperature range-storage-30 70Optimum Storage ConditionsTemperature20±5Humidity<80%RHPressure86kPa106kPaNo vapor or electromagnetic disturbances in the environment are influencing the normal working conditions of the deviceBlowing Time3 seconds breath flow≥20L/minWarm-up Time6 seconds after switching onRecovery Time1-5 minutes, depends on Ethanol Vapor ConcentrationAnalysis TimeLess than 10 secondsDisplay2.4inch three colorscreenMemory CapacityStores 6000 measured values with date and timedata can be transferred to a PC or printerCommunication Method for PrinterBluetoothPrinterMicro Thermal printer/Dot matrix PrinterDimensions(L x W x H)Approx. 140mm×60mm×24.5mmNumber of Measurements per Battery PackMore than 100 times under ambient conditionsOperation VoltageState of chargingState of operationDC5VDC 3.3 4.2VBatteries1200mAh/3.7V Lithium-ion-batteryWeightApprox. 180g Note: The device performs best under benchmark working condition. It is able to work between temperature -5 and 45 with ambient humidity from 20%RH to 90%RH. It complies with the requirements ofDOT/NHTSA in respect of measuring accuracy.